Clinical Gastroenterology Vol.16 No.2(4-2)

Theme Latest Developments in Devices Related to Colonoscopy and Progress in Insertion Techniques
Title Usefulness of Variable Stiffness Colonoscope (CF-Q240AI)
Publish Date 2001/02
Author Sumio Fujinuma Division of Endoscopy, Toho University School of Medicine Ohashi Hospital
Author Hiroaki Suda Third Department of Internal Medicine, Toho University School of Medicine
Author Satoshi Sugano Third Department of Internal Medicine, Toho University School of Medicine
Author Masanobu Furuya Third Department of Internal Medicine, Toho University School of Medicine
Author Kenichi Sakai Third Department of Internal Medicine, Toho University School of Medicine
Author Yoshihiro Sakai Division of Endoscopy, Toho University School of Medicine Ohashi Hospital
[ Summary ] A beginning endoscopist performed colonoscopies in 72 cases using a colonoscope with a flexible, adjustable shaft(CF-Q 240 Al) and 32 cases using the usual CF-Q240I. In comparison with the CF-Q240I, the CF-Q 240 Al has additional functions, by which the stiffness of the shaft can be adjusted by rotating a ring in the handle. The successful rate for insertion into the cecum, using CF-Q 240 I was 93.8%. The other hand, that for the CF-Q 240 Al was 100%, although it was 92% when using the only softer mode with the CF-Q 240 Al. The remaining 8% required the harder mode with the CF-Q 240 Al, to pass the splenic flexure. The insertion time for the CF-Q 240 Al was slightly longer than that for the CF-Q 240 I. However, the rate of these which were painless with the CF-Q 240 Al was higher than with the CF-Q 240 I. For beginning to intermediate level endoscopists the softer mode during insertion is recommended. The harder mode should be used after straightening the scope, when the scope makes a re-loop. The CF-Q 240 Al is useful, not only for expert endoscopists but also beginners.
back